Operation – KROHNE OPTIFLEX 2200 C_F EN User Manual
Page 92
6
OPERATION
92
OPTIFLEX 2200 C/F
www.krohne.com
07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
If an "Out of specification" status symbol is shown, refer to menu item 2.2.2 DIAGNOSTIC
(Configuration mode / Supervisor menu) for more data.
For data on errors, error records and error codes, refer to
Error handling
on page 93.
NE 107
symbol
shown
NE 107
Status
Description
Status
marker
shown
Error code
(Type)
Possible errors
Failure
The device does not operate
correctly. The fault message stays
on. The user cannot remove the
"Failure" message from the
Normal mode screen.
1
ERR 101 (Error)
Current Output Drift
3
ERR 102 (Error)
Temperature Out of Range
1
ERR 103 (Error)
Converter EEPROM
1
ERR 103 (Error)
Converter RAM
1
ERR 103 (Error)
Converter ROM
1
ERR 104 (Error)
Converter Voltage
2
ERR 200 (Error)
Reference Lost
2
ERR 202 (Error)
Peak Lost (Level Lost)
2
ERR 204 (Error)
Overfill
3
ERR 205 (Error)
Internal Communication
1
ERR 206 (Error)
No Probe detected
1
ERR 207 (Error)
Sensor EEPROM
1
ERR 207 (Error)
Sensor RAM
1
ERR 207 (Error)
Sensor ROM
1
ERR 208 (Error)
Oscillator Frequency
3
ERR 209 (Error)
Sensor Not compatible
2, 4
ERR 210 (Error)
Empty
Function
check
The device operates correctly, but
the measured value is incorrect.
This fault message is only
temporary. This symbol is shown
when the user configures the
device with the DTM or a HART®
Communicator.
-
-
-
Out of
specification
It is possible that the measured
value is unstable if the operating
conditions do not agree with the
device specification.
4
(Warning)
Peak Lost
4
(Warning)
Overfill
4
(Warning)
Empty
4
(Warning)
Temperature out of range