Service – KROHNE OPTIFLEX 2200 C_F EN User Manual
Page 108
![background image](/manuals/369278/108/background.png)
7
SERVICE
108
OPTIFLEX 2200 C/F
www.krohne.com
07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
7.3.4 Replacement of the OPTIFLEX 1300 signal converter
Equipment needed:
• 5 mm Allen wrench (not supplied)
• OPTIFLEX 1300 C TDR level meter
• OPTIFLEX 2200 signal converter (without process connection and probe)
• The applicable adaptor for the process connection. You can send an order for this part only or
for the OPTIFLEX 2200 signal converter with the adaptor attached. For the order code, refer
to
Order code
on page 149.
• Handbooks for all devices
• Option: Workstation (not supplied) with PACTware and DTM installed
• Option: PACTware supplement (if a workstation is used to set up and monitor the device)
• Option: HART® handheld terminal (not supplied)
Procedure 1: Record the parameters (OPTIFLEX 1300 TDR level meters)
1 Record the probe type and probe length.
2 Energize the device.
i
The device is in operation and in Normal Mode.
3 Push [>
>>>], [
] and [>>>>] to enter Configuration Mode (2.0.0 SUPERVISOR).
4 Enter the password. Push [>
>>>], ^
^
^
^],
], [
], [>>>>] and [^
^
^
^].
5 Record parameters in these menu items: A.1.4 Conversion (strapping table values), B.2.7
Probe Type, C.1.9 Blocking Distance, C.1.1.0 Tank Height, Detection Delay, C.3.1 Output Func-
tion (Output 1), C.3.2 4 mA Setting (Output1) and C.3.3 20 mA Setting (Output 1).
6 Push 4 × [^
^
^
^] to go back to the Operating Mode.
7 Push [>
>>>], 3 × [
] and [>>>>] to enter the Service menu.
8 Give the password for the SERVICE menu. If you do not have the password, speak to the sup-
plier.
9 Record parameters in these menu items: D2.1.0 Converter Offset and D.2.3.0 Mechanical Cal-
ibration Ratio.
10 Push 2 × [^
^
^
^] to go back to the Operating Mode.
11 De-energize the device.
12 Remove the electrical cables.
13 Attach the signal converter cover.
INFORMATION!
Complete the 5 procedures that follow in numerical sequence. These procedures are applicable
only to devices that were made bafore August 2009.
To get the passwords for the OPTIFLEX 1300 and OPTIFLEX 2200 Service menus, speak to the
supplier.
CAUTION!
Make sure that you also record device configuration data. This data includes basic configuration
(tank height, blocking distance etc.), output, application, display and strapping table, converter
offset and mechanical calibration ratio data. You can find this data in the Supervisor and Service
menus. If you do not have the password for the SERVICE menu, speak to the supplier.
INFORMATION!
You can see the device settings on the display screen (if your device has this option), at a
workstation with PACTware software or with a HART
®
Handheld Terminal. For more data about
the software, refer to the PACTware supplement or PACTware embedded Help for each device.