beautypg.com

Installation, 2 general information for nozzles – KROHNE OPTIFLEX 2200 C_F EN User Manual

Page 19

background image

INSTALLATION

3

19

OPTIFLEX 2200 C/F

www.krohne.com

07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en

3.5.2 General information for nozzles

CAUTION!
Follow these recommendations to make sure that the device measures correctly. They have an
effect on the performance of the device.

CAUTION!
Do not put the process connection near to the product inlet. If the product that enters the tank
touches the probe, the device will measure incorrectly.

Figure 3-5: Do not put the device near to a product inlet

1 The device is in the correct position.
2 The device is too near to the product inlet.
3 If it is not possible to put the device in the recommended position, install a deflector pipe.

Figure 3-6: How to prevent build up of product around the process connection

1 If product particles are likely to collect in holes, a nozzle is not recommended.
2 Attach the flange directly to the tank.
3 Use a threaded connection to attach the device directly to the tank.