beautypg.com

Operation – KROHNE OPTIFLEX 2200 C_F EN User Manual

Page 75

background image

OPERATION

6

75

OPTIFLEX 2200 C/F

www.krohne.com

07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en

2.0.0 Supervisor menu

Menu
No.

Function

Function description

Selection list

Default

2.1.0 COMMISSION.

2.1.0

COMMISSION.

This starts a quick set-up procedure

applicable to most applications. The

supervisor can give the tank height (TANK

HEIGHT), output function (OUTPUT

FUNC.), current output range (RANGE I),

4 mA setting (SCALE 4mA), 20 mA setting

(SCALE 20mA), error delay (ERROR

DELAY) and tag name (TAG NAME).

2.2.0 TESTS

2.2.1

SET OUTPUT

This sets the analog output to a test value

[mA] selected from a list. The output will

change to the selected value after 5

seconds, independent of the measured

value.

3.5, 4, 6, 8, 10, 12, 14, 16,

18, 20 or 22 mA

3.5 mA

2.2.2

DIAGNOSTIC

This starts the hardware test. Push [>>>>]

many times to show: the time of operation

(D1), temperature of the electronic

converter board (T1), loop current (I1),

load current (I2), voltage 5.6 V (V1), voltage

on capacitors (V2), voltage 3.3 V (V3),

amplitude reference pulse (P1), amplitude

level pulse (P2), amplitude probe end

pulse (P3), reset counter (C1). If you push

[>>>>] again, the display goes back to the

menu level.

2.3.0 BASIC PARAM.

2.3.1

TANK HEIGHT

The distance from the tank connecting

flange face / thread stop down to the tank

bottom.

min-max:

0…80 m / 0…262.48 ft

If tank height is

not specified in

customer order,

the probe length

value is used as

an alternative

2.3.2

TIME CONST.

Increasing the time constant will

smoothen the integrated readings,

decreasing will roughen the readings.

min-max:

0 to 100 seconds

5 seconds

2.3.3

PROBE LENGTH

Probe length is the distance from the

flange face / thread stop of the device

down to the bottom end of the probe

(including counterweight for cable

versions). If probe length has been

modified, enter the new value here. For

more data, refer to

How to decrease the

length of probes

on page 89.

min:

2.3.4 BLOC. DIST. +

3.1.1 COUNTERWEIGHT

max.:

40 m / 131.23 ft

This value is

given in

customer order

2.3.4

BLOC. DIST.

Blocking distance. The non-measuring

range at the top of the probe. It depends

on the probe type and the installation.

min:

0 m / 0 ft

max:

2.3.3 PROBE LENGTH

50 mm / 1.97¨

2.3.5

TAG NAME

The device has a code (tag name) to

identify it. If the tag name is given in the

customer order data, it will be set at the

factory. A maximum of 8 characters can

be used.

?

TANK 01