Operation – KROHNE OPTIFLEX 2200 C_F EN User Manual
Page 68
6
OPERATION
68
OPTIFLEX 2200 C/F
www.krohne.com
07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
Measurement definitions
Measurement name
Description
Available units
LEVEL
This is a display and an output function option. It
is the height from the bottom of the tank to the
surface of the liquid or solid contents
(Tank height - Distance).
m, cm, mm, in (inches), ft
(feet)
DISTANCE
This is a display and an output function option. It
is the distance from the face of the flange to the
surface of the liquid or solid contents of the
tank.
m, cm, mm, in (inches), ft
(feet)
CONVERSION
This is a display and an output function option. It
gives the volume or mass of the tank contents.
This data is available if you prepare a volume or
mass table in configuration mode. For data on
how to prepare the conversion table, refer to
How to configure the device to measure volume
or mass
on page 85.
kg, t, Ston, Lton, m, cm, mm,
in, ft, m3, L, gal, Imp, ft3, bbl
ULLAGE CONV.
This is a display and an output function option. It
gives the empty volume or remaining mass that
can be put in the tank. This data is available if
you prepare a volume or mass table in
configuration mode. For data on how to prepare
the conversion table, refer to
How to configure
the device to measure volume or mass
on page
85.
kg, t, Ston, Lton, m, cm, mm,
in, ft, m3, L, gal, Imp, ft3, bbl
EPSILON R
The dielectric constant of the contents of the
tank. An electrical property of the liquid or solid
contents of the tank. Also known as ε
r
, DK and
relative permittivity. This gives the strength of
the measurement pulse reflection.
This data is shown if the menu item AUTO Er
CALC (2.5.2) is used to calculate the dielectric
constant value.
OUTPUT I (mA)
The current output of the device.
mA
OUTPUT I (%)
The percentage of the current output.
0% = 4 mA. 100% = 20 mA.
%