beautypg.com

Operation – KROHNE OPTIFLEX 2200 C_F EN User Manual

Page 79

background image

OPERATION

6

79

OPTIFLEX 2200 C/F

www.krohne.com

07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en

Default values for menu items 2.4.3 SCALE 4mA and 2.4.4 SCALE 20mA

3. Service menu

4. Master menu

Probe type

SCALE 4mA

SCALE 20mA

[mm]

[inches]

[mm]

[inches]

Ø2 mm / 0.08¨ Single cable

250

9.84

L - 300

1

L - 11.81

1

Ø4 mm / 0.16¨ Single cable

200

7.87

L - 300

1

L - 11.81

1

Single rod

170

6.69

L - 300

1

L - 11.81

1

Coaxial

50

1.97

L - 200

1

L - 7.87

1

Ø4 mm / 0.16¨ double cable

100

3.94

L - 250

1

L - 9.84

1

Ø8 mm / 0.31¨ double rod

110

4.33

L - 270

1

L - 10.63

1

1 L, Probe length

Menu
No.

Function

Function description

Selection list

Default

3.0.0

SERVICE

Advanced settings. The settings in this

menu are protected with a password.

Only approved personnel can change

the parameters in this menu. For more

data, contact your local sales office.

Menu
No.

Function

Function description

Selection list

Default

4.0.0

MASTER

Factory settings. The settings in this

menu are protected with a password.

Only approved personnel can change

the parameters in this menu. For more

data, contact your local sales office.