OPERATION

6

77

OPTIFLEX 2200 C/F

www.krohne.com

07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en

2.5.5

PROD. EPS. R

Dielectric constant (ε

r

) of the the liquid or

solid contents in the tank. If possible, type

in the exact value of the product's

dielectric constant. If you do not know the

exact value, use 2.5.2 AUTO Er CALC. A

new product ε

r

that is too low will mean

that level will read higher than its true

value. This menu item is only used in TBF

mode.

min-max: 1.0 to 115.00

2.4

2.5.6

MEAS. AMP.

Measured pulse amplitude. This is the

amplitude of the signal (after reflection on

the surface of the tank contents)

compared with the amplitude of the

reference pulse. This value helps you to

set the measurement threshold in menu

item 2.5.7 MEAS. THRESH. For more data,

refer to

Thresholds and parasitic signals

on page 87.

2.5.7

MEAS. THRESH.

Measurement threshold. If it is difficult to

identify the level signal (for example: too

many parasite signals), you can increase

the threshold. This value is measured in

thousandths (1...1000). A threshold of 100

is equivalent to 10% of the amplitude of

the reference pulse at a distance of 1 m /

3.3 ft from the thread stop. For more data,

refer to

Thresholds and parasitic signals

on page 87.

min-max: 0 to 1000

60

2.5.8

PROBE END AMP

Probe end pulse amplitude. This is the

amplitude of the signal (after reflection at

the bottom of the probe) compared with

the amplitude of the reference pulse. This

value helps you to set the measurement

threshold in menu item 2.5.9 PROBE END

AMP. For more data, refer to

Thresholds

and parasitic signals

on page 87.

2.5.9

PROBE END TH.

Probe end threshold. For measurement in

TBF mode. If it is difficult to identify the

probe end signal (for example: too many

parasite signals), you can increase the

threshold of the signal. A threshold of 100

is equivalent to 10% of the amplitude of

the reference pulse at a distance of 1 m /

3.3 ft from the thread stop. For more data,

refer to

Thresholds and parasitic signals

on page 87.

min-max: 0 to 1000

50

2.6.0 COMMUNICATION

2.6.1

HART ADDRESS

Any HART

®

address greater than 0 will

activate HART

®

multidrop mode. The

current output stays constant at 4 mA. If

2.6.1 HART ADDRESS is set to 0, the device

will operate in point-to-point mode.

min-max: 0…15

0

Menu
No.

Function

Function description

Selection list

Default