6

OPERATION

94

OPTIFLEX 1300 C

www.krohne.com

03/2014 - 4000172306 - HB OPTIFLEX 1300 R07 en

• Increase the amplitude of the level threshold.

i

This value must be more than the incorrect signal. If there is no parasitic signal, we
recommend that you set the level threshold at half the amplitude of the correct signal.

• Use the same gain as that used for the device.
• Save the settings.

i

The threshold increases. It ignores the parasitic signal and uses the first signal it finds.

Figure 6-10: Threshold data on the device display screen

1 Error message "Lost" if the device cannot find a signal
2 Amplitude of the threshold (in volts)
3 Gain of the threshold
4 Distance of the monitored signal from the signal converter
5 Amplitude of the monitored signal (in volts)
6 Gain of the monitored signal

INFORMATION!
If "2" or ">2" products are selected in the Application

Application

Application

Application setup mode, then the level is the first

signal that goes across the level threshold.