OPERATION

6

81

OPTIFLEX 1300 C

www.krohne.com

03/2014 - 4000172306 - HB OPTIFLEX 1300 R07 en

C.1.13

Measuring

Mode

In Direct

Direct

Direct

Direct mode, the device measures

the time it takes to receives a reflection

of the signal from the surface of the

tank contents. In TBF

TBF

TBF

TBF mode, the device

measures the time it takes to receives

a reflection of the signal from the end

of the probe. TBF mode is used to

measure one product with a low ε

r

.

Automatic

Automatic

Automatic

Automatic mode automatically

switches the measuring mode between

Direct and TBF mode. This is sufficient

for most applications. Direct mode is

used for products with an ε

r

≥1.6 (this

depends on the probe type). If ε

r

≤1.4,

use TBF mode.

Automatic, Direct, TBF

Automatic

C.1.14

Product Er

This function is available if TBF

TBF

TBF

TBF or

Automatic

Automatic

Automatic

Automatic mode is selected in C.1.13. If

possible, type in the exact value of the

product's dielectric constant. If you do

not know the exact value, use the

default value. A new product ε

r

that is

too low will mean that level will read

higher than its true value in TBF mode.

min-max: 1.10 to 115.00

2.4

C.1.15

Gas Er

A major parameter for TDR level

measurement devices. If the gas is not

air, set the ε

r

value to the ε

r

value of the

gas.

min-max: 0.8...115

1

C.1.16

Level

Threshold

If it is difficult to identify the level

signal (for example: too many parasitic

signals), you can increase the level

threshold and the amplification (Gain)

of the signal. For more data, refer to

Thresholds and parasitic signals

on

page 92.

7 Gains, min-max: 0...+1.25 V.

Depends on

Gain (for

example, 0.5 V

Gain 6).

C.1.17

Interface

Threshold

Devices for interface applications only.

If it is difficult to identify the interface

signal (for example: too many parasite

signals), you can increase the interface

threshold and the amplification (Gain)

of the signal. For more data, refer to

Thresholds and parasitic signals

on

page 92.

7 Gains, min-max: 0...+1.25 V.

Depends on

Gain (for

example, 0.5 V

Gain 6).

C.1.18

Probe end

Threshold

Used in TBF

TBF

TBF

TBF or Automatic

Automatic

Automatic

Automatic mode. If it is

difficult to identify the probe end signal

(for example: too many parasite

signals), you can increase the probe

end threshold and the amplification

(Gain) of the signal. For more data,

refer to

Thresholds and parasitic

signals

on page 92.

7 Gains, min-max: 0...+1.25 V

or -1.25…0 V.

Depends on

Gain (for

example, 0.5 V

Gain 6).

Snapshot

Starts or stops the operation of the

snapshot signal filter. Obey the

instructions in the quick setup menu to

get snapshot data. For more data, refer

to

How to use the snapshot function to

filter parasitic signals

on page 96.

Yes, No

No

C.1.19

Units For

Tables

Sub-menu for volume and mass

conversion operations.

Table Length Unit, Conversion

Unit

C.1.19.1

Table Length

Unit

The length unit used in the conversion

table. If "free unit" is selected, the unit

device uses the unit name in menu

item C.5.1.7.

m, cm, mm, inch, ft, Free Unit

m

Menu
No.

Function

Function description

Selection list

Default