TECHNICAL DATA

130

OPTIFLEX 1300 C

www.krohne.com

03/2014 - 4000172306 - HB OPTIFLEX 1300 R07 en

8.2 Technical data

INFORMATION!

The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.

Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).

Measuring system

Measuring principle

2-wire loop-powered level transmitter; Time Domain Reflectometry (TDR)

Application range

Level measurement of liquids, pastes, slurries, powders and granulates

Primary measured value

Time between the emitted and received signal

Secondary measured value

Distance, level, volume, mass and/or interface

Design

Construction

The measurement system consists of a measuring sensor (probe) and a signal

converter which is available in a compact or remote version

Options

Integrated LCD display with sun cover (-20…+60°C / -4…+140°F);

if the ambient temperature is not in these limits, the display switches off

2nd current output

ESD protection (max. 30 kV)

Metaglas

®

(dual process sealing system for dangerous products (ammonia,

chlorine, ...))

1

Remote housing connected to the probe via a flexible conduit

Standard lengths: 2 m / 6.6 ft, 4.5 m / 14.8 ft, 9.5 m / 31.2 ft and 14.5 m / 47.6 ft

Probe end types (not for rod and coaxial probes)

Probe end types (not for rod and coaxial probes)

Probe end types (not for rod and coaxial probes)

Probe end types (not for rod and coaxial probes)

Standard: Counterweights (refer to counterweight dimensions in "Technical data:

Dimensions and weights")

Options: Turnbuckle, chuck, threaded end, crimped end, open end

Accessories

Weather protection

Max. measuring range

Double rod Ø8 mm / 0.32¨: 4 m / 13 ft

Single rod Ø8 mm / 0.32¨: 4 m / 13 ft

Single rod Ø8 mm / 0.32¨ (segmented): 6 m / 20 ft

Coaxial Ø22 mm / 0.87¨: 6 m / 20 ft

Coaxial Ø22 mm / 0.87¨ (segmented): 6 m / 20 ft

Double cable Ø4 mm / 0.16¨: 8 m / 26 ft

Single cable Ø2 mm / 0.08¨: 35 m / 115 ft (for liquids only)

Single cable Ø4 mm / 0.16¨: 35 m / 115 ft (For liquids only. An angled probe is

available on request for installations with very low ceilings or objects in the tank

that prevent installation on top of the tank.)

Single cable Ø8 mm / 0.32¨: 35 m / 115 ft (for solids only)

Tolerance, probe length

±0.5%

Dead zone

This depends on the type of probe. For more data, refer to

Measurement limits

on

page 141.

Display and User interface

Display and User interface

Display and User interface

Display and User interface

Display

LCD display

9 lines, 160×160 pixels in 8-step grayscale with 4-button keypad

Interface languages

English, German, French, Italian, Spanish, Portuguese, Japanese, Simplified

Chinese and Russian