3

INSTALLATION

26

OPTIFLEX 1300 C

www.krohne.com

03/2014 - 4000172306 - HB OPTIFLEX 1300 R07 en

• Weld a tube with an internal diameter of 12 mm / 0.5¨ to the bottom of the tank.

i

Make sure the tube aligns with the process connection at the top of the tank.

• Lower the probe into the tank.
• Put the end of the probe into the tube.

The probe counterweight has a hole with an M8 internal thread. The other probe end options are
given in the illustration.

Single rod Ø8 mm / 0.32¨

Figure 3-13: How to attach a single rod probe to keep it straight

Single cable Ø4 mm / 0.16¨

Figure 3-14: How to attach a Ø4 mm / 0.16¨ single cable probe to keep it straight

1 Probe with threaded end
2 Probe with turnbuckle
3 Probe with chuck
4 If you chose a chuck to anchor the probe, we recommend that you fit a ferrule (metal sheath - not supplied) at the

bottom of the loop to prevent cable wear

CAUTION!
If your device has a chuck, you must make the device recalculate the probe length. For the
procedure, refer to How to decrease the length of probes on page 98

. If the device does not

recalculate the probe length, it is possible that the device will not measure correctly.