Data dependencies for the 20 mA settings of outputs 1 and 2

Output function

Minimum value

Maximum value

Default

Level

>4 mA Setting for Level

Tank Height + TBO + RO

1

Tank height +

TBO - BD

2

Volume

>4 mA Setting for Volume

Max. value in the volume table

Max. value in the

volume table

Mass

>4 mA Setting for Mass

Max. value in the mass table

Max. value in the

mass table

Distance

>4 mA Setting for Distance

Tank Height + TBO + RO

1

Tank height + RO

3

Interface Level

>4 mA Setting for Interface Level

Tank height + TBO + RO

1

Tank height +

TBO - BD

2

Interface Distance

>4 mA Setting for Interface

Distance

Tank height + TBO + RO

1

Tank height + RO

3

Interface Volume

>4 mA Setting for Interface

Volume

Max. value in the volume table

Max. value in the

volume table

Interface Mass

>4 mA Setting for Interface Mass

Max. value in the mass table

Max. value in the

mass table

Ullage Volume

>4 mA Setting for Ullage Volume

Max. value in the volume table

Max. value in the

volume table

Ullage Mass

>4 mA Setting for Ullage Mass

Max. value in the mass table

Max. value in the

mass table

Layer

>4 mA Setting for Layer

Tank height + TBO + RO

1

Tank height - BD

4

Layer Volume

>4 mA Setting for Layer Volume

Max. value in the volume table

Max. value in the

volume table

Layer Mass

>4 mA Setting for Layer Mass

Max. value in the mass table

Max. value in the

mass table

1 RO = Reference Offset (C1.10). TBO = Tank Bottom Offset (C.1.11).
2 BD = Blocking Distance (C.1.9). TBO = Tank Bottom Offset (C.1.11).
3 RO = Reference Offset (C1.10)