Page 48

background image

3

INSTALLATION

48

OPTIFLEX 1300 C

www.krohne.com

03/2014 - 4000172306 - HB OPTIFLEX 1300 R07 en

3.8.6 How to install a cable probe in the tank

• Use two persons to lift the housing and the probe above the process connection.
• Hold the device 1 m / 3½ ft above the tank.
• Unwind the probe carefully into the tank.

Figure 3-38: Wind cable probes carefully

1 Do not wind cable probes less than 400 mm / 16¨ in diameter.

WARNING!
If you bend the probe too much, you will damage the device and it will not measure accurately.

Figure 3-39: Installation of devices with cable probes

1 >1 m / 3½ ft