SERVICE

7

107

OPTIFLEX 1300 C

www.krohne.com

03/2014 - 4000172306 - HB OPTIFLEX 1300 R07 en

7.3 How to replace device components

7.3.1 Service warranty

Maintenance is not necessary for most applications.

Servicing by the customer is limited by warranty to:

• The removal and installation of the signal converter housing. For more data, refer to

How to

turn or remove the signal converter

on page 53.

• The removal and installation of the complete electronic module.
• The removal and installation of the terminal module.
• The removal and installation of the HMI cover.
• Replacement of signal converters of other TDR devices:

Replacement of signal converters of other TDR devices:

Replacement of signal converters of other TDR devices:

Replacement of signal converters of other TDR devices: The removal of a BM 100 A or BM 102
signal converter and installation of the OPTIFLEX 1300 signal converter. For the procedure,
refer to

Replacement of the BM 100 signal converter

on page 114.

This housing can be detached from the flange system under process conditions.

Use only authorized service staff to repair the device.

For more data on how to prepare the device before you send it back, refer to

Returning the

device to the manufacturer

on page 124.