Operation – KROHNE OPTIFLEX 1300C EN User Manual
Page 78
6
OPERATION
78
OPTIFLEX 1300 C
www.krohne.com
03/2014 - 4000172306 - HB OPTIFLEX 1300 R07 en
B. Test
A.6 Quick Link 5
Quick Link 5
Direct link to an item in the
Advanced Setup menu
Go to a function in the
Advanced Setup menu and
push [>>>>] for 1 second. You
can store up to 5 functions
in this way.
Display
Mode
1 Only "Air Gap" is available for solid applications
2 Only "1" is available for solid applications
3 Only the "Level" menu parameter is available for tanks with one product. If there are 2 or more products in the tank and there is no air
gap, only "Interface" is available in this list (the device measures the top interface only). If the device has the "Interface" option, only
"Interface" and "Level + Interface" are available in this list.
4 This step is ignored if you use the "Complete" setup mode
5 This value agrees with the customer specification or probe length
6 Optional
Menu
No.
Function
Function description
Selection list
Default
B.1 Test
B.1.1
Show Output 1
This displays analogue output 1 value
[mA].
Read only
B.1.2
Set Output 1
This sets analogue output 1 to a test
value [mA] selected from a list. Output
will change to the selected value,
independent of the measured value.
3.6, 4, 6, 8, 10, 12, 14, 16, 18, 20 or
22 mA
4 mA
B.1.3
Show Output 2
This displays analogue output 2 value
[mA].
Read only
B.1.4
Set Output 2
This sets analogue output 2 to a test
value [mA] selected from a list. Output
will change to the selected value,
independent of the measured value.
3.6, 4, 6, 8, 10, 12, 14, 16, 18, 20 or
22 mA
4 mA
B.1.5
Internal Test
This initiates the hardware test. The
device displays the results.
Read only
B.2 Information
B.2.1
Outputs
Analogue output settings. This includes
assigned functions, 4… 20 mA scale
settings, error handling and HART
®
parameters.
Read only
B.2.2
15 Min Log
A log of output values for the last 15
minutes. A log is taken every 10
seconds and displayed on a graph.
Read only
B.2.3
Device ID
This displays device order no, V-no,
service no, Ex approval, Main CPU
version and Companion CPU version.
Read only
B.2.4
Quick Setup
Sum.
A summary of the parameters entered
in the quick setup menu.
Read only
B.2.5
TAG Number
The TAG number can be seen and
updated here.
?
TAGNO0123
4567890
Menu
No.
Step
Function
Function description
Selection list
Default