3.8.3 How to assemble the segmented coaxial probe

CAUTION!
Make sure that the length of the probe is correct. If the probe is too long, refer to How to
decrease the length of probes on page 98

.

Figure 3-33: Equipment needed to assemble the coaxial probe

1 Converter and process connection
2 HC M4×20 screws (1 screw per probe segment)
3 Lock washers (1 pair of washers per probe segment)
4 Top (quantity: 1), middle (quantity: 1 or more) and bottom (quantity: 1) segments of the signal rod
5 PTFE spacer (1 spacer per probe segment)
6 Middle (quantity: 1 or more) and bottom (quantity: 1) segments of the coaxial tube
7 Union nut with 2 socket set screws M5×5 (1 union nut per segment of the coaxial tube)
8 Tools: two 7 mm open-end wrenches (not supplied)
9 Tools: two pipe (Stillson) wrenches (not supplied)
10 Tools: one 2.5 mm Allen wrench and one 2 mm Allen wrench (not supplied)