8.5 Measurement limits
Measurement limits in mm and inches
80 is ε
r
of water; 2.3 is ε
r
of oil
Double cable and double rod probes
Figure 8-13: Measurement limits of the double cable probes (on the left side) and double rod probes (on the right side)
1 A1, Top dead zone:
A1, Top dead zone:
A1, Top dead zone:
A1, Top dead zone: Distance from the flange to the top limit of the measuring range. Refer to the notes and table that
follow.
2 A2, Bottom dead zone:
A2, Bottom dead zone:
A2, Bottom dead zone:
A2, Bottom dead zone: Length at the end of the probe, where measurement is not linear.
3 D, non measurement zone:
D, non measurement zone:
D, non measurement zone:
D, non measurement zone: Zone where measurement cannot be taken.
4 Product 1
Product 1
Product 1
Product 1
5 Gas (Air)
Gas (Air)
Gas (Air)
Gas (Air)
6 L, Probe length:
L, Probe length:
L, Probe length:
L, Probe length: Length specified by the customer in the order. This is also the maximum measuring length for some
probe types in direct mode and all devices that operate in TBF mode.
7 Tank Height
Tank Height
Tank Height
Tank Height
INFORMATION!
h
hh
h is the height of the nozzle. dddd is the diameter of the tank nozzle.
INFORMATION!
•
If h < d, then the top dead zone (A1) is equal to the top dead zone for the probe only. Refer to
the table that follows.
•
If h ≥ d, then the top dead zone (A1) is equal to the tank nozzle height plus the top dead zone
for the probe.
Probes
Top dead zone, A1
ε
r
= 80
Bottom dead
zone, A2
ε
r
= 80
Top dead zone, A1
ε
r
= 2.3
Bottom dead
zone, A2
ε
r
= 2.3
[mm]
[inches]
[mm]
[inches]
[mm]
[inches]
[mm]
[inches]
Double rod
Double cable
125
4.9
10
0.4
165
6.5
50
1.95