6.4 Further information on device configuration
6.4.1 Protection of the device settings
The Passwords
Passwords
Passwords
Passwords menu lets you change the supervisor password.
How to change the supervisor password
• Go to Supervisor > Advanced setup > Device setup > Passwords > Supervisor
Supervisor > Advanced setup > Device setup > Passwords > Supervisor
Supervisor > Advanced setup > Device setup > Passwords > Supervisor
Supervisor > Advanced setup > Device setup > Passwords > Supervisor.
• Type in the new 6-character password (push the 4 buttons in any sequence).
• Type in the new 6-character password again.
i
If the second entry is not the same as the first, the device will display the error message
"Password mismatch". Push [>>>>] and[
] at the same time and type in the new 6-character
password again.
• Push [>>>>] and [
] (Esc) at the same time to go back to the "save settings" window.
• Select Save
Save
Save
Save and push [^
^
^
^].
i
The device will go back to operator mode.
6.4.2 Network configuration
The device uses HART
®
communication to send information to HART
®
-compatible equipment. It
can operate in either point-to-point or multidrop mode. The device will communicate in
multidrop mode if you change the HART
®
address of output 1.
How to change from point-to-point to multidrop mode
• Enter supervisor mode.
• Go to Advanced Setup > Output 1 (HART) > HART Address
Advanced Setup > Output 1 (HART) > HART Address
Advanced Setup > Output 1 (HART) > HART Address
Advanced Setup > Output 1 (HART) > HART Address.
• Enter a value between 1 and 15 and push [^
^
^
^] to confirm (see the caution below).
• Push the Escape buttons ([>>>>] + [
]) until you get the save/cancel screen.
• Select save.
• Push [^
^
^
^].
i
Output 1 switches to multidrop mode. The current output is set to 4 mA. This value does not
change in multidrop mode.
INFORMATION!
Make a note of the password and keep it in a safe place. If you lose the password, please contact
your supplier.
INFORMATION!
For more data, refer to Networks on page 59
.
CAUTION!
Make sure that the address for this device is different from others in the multidrop network.