Technical data – KROHNE OPTIFLEX 1300C EN User Manual
Page 132
8
TECHNICAL DATA
132
OPTIFLEX 1300 C
www.krohne.com
03/2014 - 4000172306 - HB OPTIFLEX 1300 R07 en
Installation conditions
Process connection size
Refer to "Installation: How to prepare the tank before you install the device" and
"Technical data: Measurement limits"
Process connection position
Make sure that there are not any obstructions directly below the process
connection for the device.
Dimensions and weights
Refer to "Technical data: Dimensions and weights"
Material
Housing
Standard: Polyester-coated aluminium
Option: Stainless steel (1.4404 / 316L)
3
Single rod (single-piece)
Standard: Stainless steel (1.4404 / 316L)
Option: Stainless steel (1.4404 / 316L) in a PVDF protective sheath, Hastelloy
®
C-22
(2.4602)
4
On request: Stainless steel (1.4404 / 316L) in a PVC or PP protective sheath
On request: Monel; Tantalum; Titanium; Duplex
Single rod (segmented)
Standard: Stainless steel (1.4404 / 316L)
Double rod
Standard: Stainless steel (1.4404 / 316L)
Option: Hastelloy
®
C-22 (2.4602)
On request: Monel; Tantalum; Titanium; Duplex
Coaxial (single-piece)
Standard: Stainless steel (1.4404 / 316L)
Option: Hastelloy
®
C-22 (2.4602)
Coaxial (segmented)
Standard: Stainless steel (1.4404 / 316L)
Single cable
Standard: Stainless steel (1.4401 / 316)
Option: Hastelloy
®
C-22 (2.4602)
– only for the Ø2 mm / 0.08¨ single cable probe
On request: FEP-coated stainless steel (-20...+150°C / -4...+302°F)
– only for the Ø4 mm / 0.16¨ single cable probe
Double cable
Stainless steel (1.4401 / 316)
Process fitting
Standard: Stainless steel (1.4404 / 316L)
Option: Hastelloy
®
C-22 (2.4602)
On request: Monel; Tantalum; Titanium; Duplex
Gaskets
FKM/FPM (-40…+200°C / -40…+392°F); Kalrez
®
6375 (-20…+200°C / -4…+392°F);
EPDM (-50...+150°C / -58...+302°F) – all probes except single cable Ø8 mm / 0.32¨
5
Weather protection (Option)
Stainless steel (1.4301 / 304)
Protective sheath
(On request for single rod only)
PP (-40…+90°C / -40…+194°F); PVC (-15…+80°C / +5…+176°F);
PVDF (-40…+150°C / -40…+302°F)
Conduit for remote housing
(Option)
Zinc-coated steel in a PVC sheath (-40...+105°C / -40...+221°F)
Process connections
Thread
Thread
Thread
Thread
Single cable Ø2 mm / 0.08¨
G ½ (ISO 228); ½ NPT (ASME B1.20.1); ½ NPTF (ASME B1.20.3 – for the HT/HP
version)
Single cable Ø8 mm / 0.32¨
Double cable Ø4 mm / 0.16¨
Double rod Ø8 mm / 0.32¨
G 1½ (ISO 228); 1½ NPT (ASME B1.20.1)
All other probes