beautypg.com

AMETEK Lx Series II User Manual

Page 86

background image

User Manual

Lx / Ls Series II

82

ENTRY

DESCRIPTION

BUS

Bus mode. The transient system is started by a bus
command or a group execute trigger (GET).

EXT

External mode. The transient system is started by a
user-provided external TTL trigger signal on
TRIGGER IN1.

TRAN STEP

Indicates the transient system execution mode. Available modes
are:

AUTO

When triggered, the transient system will
automatically execute each list point sequentially
without waiting for a trigger between list points. This
execution is paced by the dwell time set for each
data point.

ONCE

When triggered, the transient system will execute
the first list point and wait for a new trigger once the
dwell time expires. This allows triggered execution of
each step in the transient list.

List parameters:

VOLT

Step #

Voltage set point

VSLEW

Step #

Voltage slew rate in V/sec

FREQ

Step #

Frequency set point

FSLEW

Step #

Frequency slew rate in Hz/sec

FUNC

Step #

Waveform selection. Available choices are Sine,
Square, Clipped or any of the user provided
waveforms in waveform memory.

PAUSE

Step #

Delay

CURR

Step #

Current set point

PHASE

Step #

Phase angle set point. (Not relevant for phase A if
clock mode is internal.)

DWELL

Step #

Dwell time in seconds.
Range is 0.001 to 900000

TTLT

Step #

ON:

Generates an output trigger pulse at this list

step.
OFF:

No output trigger.

The output trigger is available on the TRIG OUT1
SMC connector on the rear panel.