Lx / Ls Series II
82
ENTRY
DESCRIPTION
BUS
Bus mode. The transient system is started by a bus
command or a group execute trigger (GET).
EXT
External mode. The transient system is started by a
user-provided external TTL trigger signal on
TRIGGER IN1.
TRAN STEP
Indicates the transient system execution mode. Available modes
are:
AUTO
When triggered, the transient system will
automatically execute each list point sequentially
without waiting for a trigger between list points. This
execution is paced by the dwell time set for each
data point.
ONCE
When triggered, the transient system will execute
the first list point and wait for a new trigger once the
dwell time expires. This allows triggered execution of
each step in the transient list.
List parameters:
VOLT
Step #
Voltage set point
VSLEW
Step #
Voltage slew rate in V/sec
FREQ
Step #
Frequency set point
FSLEW
Step #
Frequency slew rate in Hz/sec
FUNC
Step #
Waveform selection. Available choices are Sine,
Square, Clipped or any of the user provided
waveforms in waveform memory.
PAUSE
Step #
Delay
CURR
Step #
Current set point
PHASE
Step #
Phase angle set point. (Not relevant for phase A if
clock mode is internal.)
DWELL
Step #
Dwell time in seconds.
Range is 0.001 to 900000
TTLT
Step #
ON:
Generates an output trigger pulse at this list
step.
OFF:
No output trigger.
The output trigger is available on the TRIG OUT1
SMC connector on the rear panel.