Number

Message String

Cause

Remedy

test

43

"Current self test error,
output 1

No. 1/A amplifier in
Aux. Source has no
output during Self-
test.

Contact AMETEK service department
at

[email protected]

44

"Current self test error,
output 2

No. 2/B amplifier in
Aux. Source has no
output during Self-
test.

Contact AMETEK service department
at

[email protected]

45

"Current self test error,
output 3

No. 3/C amplifier in
Aux. Source has no
output during Self-
test.

Contact AMETEK service department
at

[email protected]

216

" RS-232 receiver
framing error"

Communication
failure.

Check RS232 port settings and cable.

217

" RS-232 receiver parity
error"

Communication
failure.

Check RS232 port settings and cable.

218

" RS-232 receiver
overrun error"

Communication
failure.

Check RS232 port settings and cable.

402

"CAL password is
incorrect"

Calibration password
does not equal high
voltage range value.

Re-enter correct password.

403

"CAL not enabled"

No password entered
for calibration

Enter correct CAL password.

600

"Systems in mode:list
have different list
lengths"

Transient lists have
unequal lengths

Check list settings and correct to same
no of data points.

601

"Requested voltage and
waveform exceeds peak
voltage capability"

Wave shape selected
and RMS voltage
combine to exceed
peak voltage
capability.

Reduce RMS or crest factor of wave
shape.

602

"Requested voltage and
waveform exceeds
transformer volt-second
rating"

The selected wave
shape exceeds
output transformer
capability.

The volt-second product of he wave
form (magnitude and time in the + and
– half of wave form).

603

"Command only applies
to RS-232 interface"

Command not
relevant for GPIB
interface.

Do not use command.

604

"Trigger received before
requested number of
pre-trigger readings"

Data acquisition pre-
trigger buffer not
filled yet.

Hold off trigger or reduce pre-trigger
delay.

605

"Requested RMS current
too high for voltage
range"

Max RMS current is
function of voltage
range selected.

Reduce programmed RMS current
limit or select low voltage range.

606

"Waveform data not
defined"

No waveform name
specified

Specify waveform name before
sending waveform data.

607

"VOLT,VOLT:SLEW, and
FUNC:SHAPe modes
incompatible"

Conflict between
wave shape and
programmed slew

Reduce slew or change waveform
type.

608

"Measurement
overrange"

Measurement data
out of range.

609

"Output buffer overrun"

Too much data in

Check receive mode on application