beautypg.com

AMETEK Lx Series II User Manual

Page 220

background image

User Manual

Lx / Ls Series II

216

Number

Message String

Cause

Remedy

-311

"Memory error"

Waveform memory
checksum error.

May be the result of incomplete user-
defined waveform download. Check
interface and try downloading
waveform again. Successful download
may clear this error condition.

Alternatively, use TRAC:DEL:ALL
command to clear waveform memory.

-314

"Save/recall memory
lost"

User setup register
contents lost

Store setup in same register again.

-315

"Configuration memory
lost"

Hardware
configuration settings
lost.

Contact AMETEK service department
at

[email protected]

to obtain instructions on restoring
configuration data.

-330

"Self-test failed"

Internal error

Contact AMETEK service department
at

[email protected]

-350

"Queue overflow"

Message queue full.

Too many message. Read status
using SYST:ERR query until 0, "No
Error" is received indicating queue
empty.

-400

"Query error"

Unable to complete
query.

Check programming manual for
correct query format and parameters

-410

"Query INTERRUPTED"

Query issued but
response not read.

Check application program for correct
flow. Response must be read after
each query to avoid this error.

-420

"Query
UNTERMINATED"

Query incomplete.

Check for terminator after query
command.

-430

"Query DEADLOCKED"

Query cannot be
completed

Check application program for multiple
queries

-440

"Query
UNTERMINATED"

Query incomplete.

Check for terminator after query
command.

0

"No error"

No errors in queue

2

" Non-volatile RAM
CONFIG section
checksum failed"

Controller failure
during Self-test.

Contact AMETEK service department
at

[email protected]

3

" Non-volatile RAM CAL
section checksum failed"

Controller failure
during Self-test.

Contact AMETEK service department
at

[email protected]

4

" Non-volatile RAM
WAVEFORM section
checksum failed"

Controller failure
during Self-test.

Contact AMETEK service department
at

[email protected]

10

"Ram self test

Controller failure
during Self-test.

Contact AMETEK service department
at

[email protected]

40

"Voltage self test error,
output 1

No. 1/A amplifier in
Master source has no
output during Self-
test.

Contact AMETEK service department
at

[email protected]

41

"Voltage self test error,
output 2

No. 2/B amplifier in
Master source has no
output during Self-
test

Contact AMETEK service department
at

[email protected]

42

"Voltage self test error,
output 3

No. 3/C amplifier in
Master source has no
output during Self-

Contact AMETEK service department
at

[email protected]

This manual is related to the following products: