2.7.2 EHV Option Ls Series -Supplemental Specifications

Specifications for Ls units with -EHV voltage range option installed are same the Lx with

–EHV

option. See paragraph 2.6.2.

2.7.3 -HF Option Supplemental Specifications

Specification for Ls units with -HF frequency range option installed are same as standard Ls unit
except where noted below.

Frequency Range

Model

3000Ls-HF, 4500Ls-HF

, 6000Ls-HF

9000Ls/2-HF, 13500Ls/3-HF

12000Ls/3-HF, 18000Ls/3-HF

Three phase mode

45 Hz - 5000 Hz

45 Hz

– 2000

1

Hz

Single phase mode

45 Hz - 5000 Hz

45 Hz - 2000

1

Hz

Output

Programming
Accuracy:

< 1000 Hz

± (0.05% + 0.25 V) from 5V to FS.

> 1000 Hz

± (0.1% + 0.1% / KHz + 0.3V) from 5V to FS

Output Noise:
(20 kHz to 1 MHz)

< 250 mV

RMS typical

For output frequencies up to 1000 Hz, refer to standard measurement specifications. For output
frequencies above 1000 Hz, see table below.

Parameter

Range

Accuracy ( )

Resolution

1 Phase Mode

3 Phase Mode

Frequency

2

45.00 - 5000.0 Hz

0.1% ± 1 digit

0.01 Hz to 81.91 Hz
0.1 Hz to 819.1 Hz
1 Hz > 819.1 Hz

Phase

45.00 - 100.0 Hz
100.0 - 2000 Hz
> 2000 Hz

0.5°


0. 1°

RMS Voltage

0 - 400 Volts
< 1000 Hz
> 1000 Hz

0.05% + 0.25V

0.1% + 0.1% / KHz + 0.3V

0.01 Volt

RMS Current

0 - 50 Amps

0.5% + 0.15A

0.5% + 0.05A

0.001 Amp

Peak Current

0 - 150 Amps

0.5% + 0. 5A

0.5% + 0.15A

0.001 Amp

Crest Factor

1.00

– 10.00

1.5 %

1.5 %

0.01

VA Power

0 - 6 KVA

0.5% + 9 VA

0.5% + 3 VA

1 VA

Real Power

0 - 6 KW

0.5% + 9 W

0.5% + 3 W

1 W

Power Factor

0.00 - 1.00

0.03

0.01

0.01

1

Multi box systems

–HF frequency limit is set to 5000 Hz but performance to specification is guaranteed only to 2000 Hz. Above

2000 Hz, full power may not be available due to amplifier unbalance. Current measurements accuracy is affected by system
interface cable cross talk above 2000 Hz.

2

Frequency measurement specifications valid with output voltage of 30Vrms or higher. If output relay is open,

frequency measurement will return 0.0 Hz.