AMETEK Lx Series II User Manual

Page 44

background image

User Manual

Lx / Ls Series II

40

Parameter

Supplemental Specification

(harmonics and noise to 300 kHz)

Frequency

Range Standard Lx:

45

– 819 Hz

With

–HF option:

45 - 5000 Hz.

Resolution Standard Lx:

0.1 Hz

With

–HF option:

1 Hz

Accuracy ± 0.025%

Phase

Phase Resolution Standard:

standard specifications apply.

With

–HF option:

0.5°

Phase Accuracy Standard:

standard specifications apply.

With

–HF option:

< 2° + 1°/kHz

Note: Specification valid under equal load conditions.

2.7.6

–EXS Option Ls Series - Supplemental Specifications

The

–EXS (External Sync) option allows the output frequency of the AC source to be synchronized to an

external TTL level clock signal. Note that some implementation changes were made starting with firmware
revision 1.50 as indicated below.

The following supplemental specifications apply when the Ls is configured with the external sync option. (-
EXS).

Parameter

Supplemental Specification

Input

Voltage Input TTL Level square wave.

Impedance 10 KOhm.

Frequency

Range Same as internal clock mode. See configuration limits.

Max Sync Input Slew Rate Units with firmware revision less than 1.5: < 80 Hz / sec.

Units with firmware revision 1.5 or higher: < 250 Hz / sec

Max Sync Step Units with firmware revision less than 1.5: < 20 Hz.

Units with firmware revision 1.5 or higher: < 200 Hz

Errors

Error 804

External Sync Error

Following conditions will result in Error 804:
1.Loss of sync input signal
2. Sudden changes in sync input frequency (see max. sync step).
3. Fast changes in sync input frequency (see max sync slew rate).
Response time to detect sync fault is 400 msec. If a sync error occurs, the
output relay will be opened. Units with firmware revision less than 1.5 will
remain in External sync mode. Units with firmware revision 1.5 or higher
will revert back to Internal Sync mode.

Mode Selection

Units with firmware revision less than 1.5:
When switching between INT and EXT sync mode, the output of the AC
source will be dropped momentarily.
Units with firmware revision 1.5 or higher:
When switching form INT to EXT sync mode, the internal frequency is set
to the detected ext sync frequency first, then the mode is switched to
external sync. This sync process takes about 200 msec so there will be this
much delay between receiving the EXT sync command and the actual
change of sync mode. The output does not drop out during this time.

This manual is related to the following products: