AMETEK Lx Series II User Manual
Page 173

User Manual
Lx / Ls Series II
169
TAC 105 Tests
– Revision E, F
TAC105
Frequency Modulation Rev E, F
Condition
Duration (s)
A
Fmod F = 1 Hz/sec, Fmod Amp = 4 Hz± 2 Hz
1800
B
Fmod F = 5 Hz/sec, Fmod Amp = 4 Hz± 2 Hz
1800
C
Fmod F = 10 Hz/sec, Fmod Amp = 4 Hz± 2 Hz
1800
D
Fmod F = 25 Hz/sec, Fmod Amp = 4 Hz± 2 Hz
1800
E
Fmod F = 100 Hz/sec, Fmod Amp = 4 Hz± 2 Hz
1800
TAC 105 Tests
– Revision B, C, D
TAC105
Frequency Modulation Rev B, C, D
Condition
Duration (s)
A
Fmod F = 1 Hz/sec, Fmod Amp = ±5 Hz
1800
B
Fmod F = 5 Hz/sec, Fmod Amp = ±1.75 Hz
1800
C
Fmod F = 10 Hz/sec, Fmod Amp = ±1.20 Hz
1800
D
Fmod F = 25 Hz/sec, Fmod Amp = ±0.85 Hz
1800
E
Fmod F = 100 Hz/sec, Fmod Amp = ±0.58 Hz
1800
TAC 105 Tests
– Revision A
TAC105
Frequency Modulation Rev A
Condition
Duration (s)
A
Fmod F = 1 Hz/sec, Fmod Amp = ±4 Hz
1800
B
Fmod F = 5 Hz/sec, Fmod Amp = ±4 Hz
1800
C
Fmod F = 10 Hz/sec, Fmod Amp = ±4 Hz
1800
D
Fmod F = 25 Hz/sec, Fmod Amp = ±4 Hz
1800