AMETEK Lx Series II User Manual
Page 139

User Manual
Lx / Ls Series II
135
VOLT FREQ MAX
This test will set the voltage and frequency to a level defined by Table 8-9. The test will last for 30
minutes. The test will be repeated using the voltage from Table 8-8and frequency from Table
8-9. The
key (backspace) will terminate the test at any time.
VOLT UNBALANCE
Standard/Group
RTCA
A(CF)
A(NF)
A(WF)
Voltage offset
8
8
8
10
Frequency
400
360/440
360/650
360/800
Standard/Group
Group1
Group2
Group3
Voltage offset
8
8
12
Frequency
400
360/650
360/800
Table 8-10: Emergency Voltage Unbalance
This test will change the output voltage for phase A and B to 122V and phase C to a voltage
lower by a value specified by an offset. Refer to Table 8-10 for the offset value and the
Frequency. The test will repeat with the same frequency and phase A and B volt is set to 100V
and phase C set to a higher voltage specified by the offset. The test will last 30 minutes. The
test will be repeated for a second Frequency if applicable. The test can be terminated at any
time.
The
key (backspace) will terminate the test at any time.