4.8 Setting the Power-on Initialization Values
The power source is shipped with default factory settings when the unit is powered up. The
factory settings are:
Parameter
Factory default setting
Voltage
0.0 Volt
Frequency
60 Hz
Waveform
SINE
Current limit
Maximum available current.
Current Protection Mode
Constant Current mode
Output state
OFF
Phase Rotation
A = 0°, B = 240°, C = 120°
Local / Remote State
Local. Front panel unlocked.
Table 4-3: Factory Default Power on Settings
It is possible to change the power on initialization values in one of two ways:
1. Using the IEEE-488, RS232C, LAN or USB bus interface.
2. Using the front panel.
To change the power on initialization values from the front panel, proceed as follows:
1. Set the AC power source output parameters from the front panel as you want to power
up the unit.
2. Save this setting to setup register 0 from the REGISTERS menu.
3. Select the CONFIGURATION menu and move to the POWER ON field.
4. Change the POWER ON field to REG0.
5. This will recall the settings contained in register 0 at power up.