AMETEK Lx Series II User Manual
Page 187

User Manual
Lx / Ls Series II
183
SVF 106 Tests
Voltage distortion spectrum test require the use of additional external equipment. In particular, a
high bandwidth (100Khz) AC source, a high frequency coupling transformer and coupling
network as shown in figure SVF106-1 of the Mil-Std 704 standard. For this test, the Lx/Ls Series
AC source can be programmed to a steady state condition of 360, 400, 600 and 800 Hz and
Vnominal of 115 Vrms or 230 Vrms.
Figure 11-10: Required SVF-106 Test Setup.
SVF 107 Tests
– Revision F
SVF107
Total Voltage Distortion Rev F
Condition
Duration (s)
Run at 400, Repeat at 360, 600 and 800 Hz
A
0.05 % VTHD
1800
F = 100%
H3 = 2.75%
H5 = 2.75%
H7 = 1.97%
H9 = 1.53%
H11 = 1.25%
H13 = 1.06%
H15 = 0.92%
SVF 108 Tests
SVF108 are DC offset tests and are not supported by the Lx/Ls Series AC Power sources.
Contact AMETEK Programmable Power for DC Supply information.
(