SVF 106 Tests

Voltage distortion spectrum test require the use of additional external equipment. In particular, a
high bandwidth (100Khz) AC source, a high frequency coupling transformer and coupling
network as shown in figure SVF106-1 of the Mil-Std 704 standard. For this test, the Lx/Ls Series
AC source can be programmed to a steady state condition of 360, 400, 600 and 800 Hz and
Vnominal of 115 Vrms or 230 Vrms.

Figure 11-10: Required SVF-106 Test Setup.

SVF 107 Tests

– Revision F

SVF107

Total Voltage Distortion Rev F

Condition

Duration (s)

Run at 400, Repeat at 360, 600 and 800 Hz

A

0.05 % VTHD

1800

F = 100%

H3 = 2.75%

H5 = 2.75%

H7 = 1.97%

H9 = 1.53%

H11 = 1.25%

H13 = 1.06%

H15 = 0.92%

SVF 108 Tests

SVF108 are DC offset tests and are not supported by the Lx/Ls Series AC Power sources.
Contact AMETEK Programmable Power for DC Supply information.
(