Lx / Ls Series II

218

Number

Message String

Cause

Remedy

output buffer.

program. Program is not reading data
sent by AC source.

610

"Command cannot be
given with present
SYST:CONF setting"

Command conflicts
with available
hardware or firmware
option settings.

Check configuration for available
options and features.

801

"Output volt fault"

- Output voltage does
not match program
value when ALC is
on.
- Over load
- Voltage kick-back
- No output voltage

Load exceeds current limit and unit is
in Constant Voltage (CV) mode of
operation.
- Reduce load or increase CL setting.
Output voltage is driven above
programmed voltage by external
influence (Load, voltage kickback, etc.)

802

"Current limit fault"

Current limit
exceeded.

Load exceeds current limit and unit is
in Constant Voltage (CV) mode of
operation. Reduce load or increase CL
setting

803

"Temperature fault"

Amplifier heat sink
temp. too high.

Reduce load. Ensure proper air flow
and exhaust clearance. Check fan(s)
for operation.

804

"External sync error"

Could not sync to
external sync signal.

External sync signal missing,
disconnected or out of range.

805

"Initial memory lost"

Initial settings could
not be recalled at
power-up.

Save power on settings again to
overwrite old content.

806

"Limit memory lost"

Hardware
configuration settings
could not be recalled
at power-up.

Contact AMETEK service department
at

[email protected]

to obtain instructions on restoring
configuration data.

807

"System memory lost"

Memory corrupted
during power-up.

Recycle power.

808

"Calibration memory lost" Calibration data lost

during power-up.

Contact AMETEK service department
at

[email protected]

to obtain instructions on restoring
calibration data or recalibrate unit.

813

"Missing list parameter"

One or more
transient list
parameters missing.

Check programmed lists.

814

"Voltage peak error "

Peak voltage
exceeds internal bus
voltage

This error may occur when selecting
user defined wave shapes with higher
crest factors. Reduce programmed
RMS value.

815

"Slew time exceed dwell" Time needed to slew

to final value is less
than dwell time.

Check dwell times in transient list
settings. Increase dwell time or
change slew rate for affected
parameter.

816

"Illegal during transient"

Operation requested
not available while
transient is running.

Wait till transient execution is
completed or abort transient execution
first.

817

"Output relay must be
closed"

Transient
programmed with
output relay open.

Close relay before attempting transient
operation.

819

"Clock and sync must be
internal"

Operation not
possible with external

Switch to internal sync. (Default)