AMETEK Lx Series II User Manual
Page 172

User Manual
Lx / Ls Series II
168
11.7.2 TAC Group Tests
Three phase 400 Hz AC tests
TAC 102 Tests - Revisions A, B, C, D, E, F
TAC102
Steady State Voltage and Frequency - Rev A through F
Condition
Duration (s)
Balanced Voltage
Nominal Voltage
A
Vnom=115V, Fnom=400 Hz - Nominal
1800
B
Vnom=115V, Flow=393 Hz - Nominal
1800
C
Vnom=115V, Fhigh=407 Hz - Nominal
1800
Low Voltage
D
Vlow =108V, Fnom=400 Hz - Low
1800
E
Vlow=108V, Flow= 393 Hz- Low
1800
F
Vlow=108V, Fhigh=407 Hz - Low
1800
High Voltage
G
Vhigh=118V, Fnom=400 Hz - High
1800
H
Vhigh=118V, Flow=393 Hz- High
1800
I
Vhigh=118V, Fhigh=407 Hz - High
1800
Unbalanced Voltage
J
Va, Vb+3V, Vc+3V, Fnom
1800
K
Va, Vb-3V, Vc-3V, Fnom
1800
TAC 103 Tests - Revisions A, B, C, D, E, F
TAC103
Voltage Phase Difference - Rev A through F
Condition
Duration (s)
A
A = 0, B = 116, C = 240
1800
B
A = 0, B = 124, C = 240
1800
TAC 104 Tests - Revisions B, C, D, E, F
TAC104
Voltage Modulation - Rev B through F
Condition
Duration (s)
A
Vmod F = 1.0 Hz, V = 0.375 Vrms
1800
B
Vmod F = 1.7 Hz, V = 0.375 Vrms
1800
C
Vmod F = 10 Hz, V = 2.5 Vrms
1800
D
Vmod F = 25 Hz, V = 2.5 Vrms
1800
E
Vmod F = 70 Hz, V = 0.375 Vrms
1800
F
Vmod F = 100 Hz, V = 0.375 Vrms
1800
G
Vmod F = 200 Hz, V = 0.375 Vrms
1800
TAC 104 Tests
– Revision A
TAC104
Voltage Modulation - Rev A
Condition
Duration (s)
A
Vmod F = 1.0 Hz, V = 0.5 Vpp
1800
B
Vmod F = 1.7 Hz, V = 0.5 Vpp
1800
C
Vmod F = 10 Hz, V = 3.5 Vpp
1800
D
Vmod F = 25 Hz, V = 3.5 Vpp
1800
E
Vmod F = 70 Hz, V = 0.5 Vpp
1800
F
Vmod F = 100 Hz, V = 0.5 Vpp
1800
G
Vmod F = 200 Hz, V = 0.5 Vpp
1800