AMETEK Lx Series II User Manual
Page 160

User Manual
Lx / Ls Series II
156
SAC102
Steady State Voltage and Frequency - Rev A
Condition
Duration (s)
High Voltage
G
Vhigh=118V, Fnom=400 Hz - High
1800
H
Vhigh=118V, Flow=380 Hz- High
1800
I
Vhigh=118V, Fhigh=420 Hz - High
1800
SAC 103 Tests
Voltage Phase Difference tests do not apply for single-
phase EUT‟s.
SAC 104 Tests - Revisions B, C, D, E, F
SAC104
Voltage Modulation Rev B - F
Condition
Duration (s)
A
Vmod F = 1.0 Hz, V = 0.375 Vrms
1800
B
Vmod F = 1.7 Hz, V = 0.375 Vrms
1800
C
Vmod F = 10 Hz, V = 2.5 Vrms
1800
D
Vmod F = 25 Hz, V = 2.5 Vrms
1800
E
Vmod F = 70 Hz, V = 0.375 Vrms
1800
F
Vmod F = 100 Hz, V = 0.375 Vrms
1800
G
Vmod F = 200 Hz, V = 0.375 Vrms
1800
SAC 104 Tests - Revisions A
SAC104
Voltage Modulation Rev A
Condition
Duration (s)
A
Vmod F = 1.0 Hz, V = 0.5 Vpp
1800
B
Vmod F = 1.7 Hz, V = 0.5 Vpp
1800
C
Vmod F = 10 Hz, V = 3.5 Vpp
1800
D
Vmod F = 25 Hz, V = 3.5 Vpp
1800
E
Vmod F = 70 Hz, V = 0.5 Vpp
1800
F
Vmod F = 100 Hz, V = 0.5 Vpp
1800
G
Vmod F = 200 Hz, V = 0.5 Vpp
1800
SAC 105 Tests
– Revisions E, F
SAC105
Frequency Modulation Rev E, F
Condition
Duration (s)
A
Fmod F = 1 Hz/sec, Fmod Amp = 4 Hz± 2 Hz
1800
B
Fmod F = 5 Hz/sec, Fmod Amp = 4 Hz± 2 Hz
1800
C
Fmod F = 10 Hz/sec, Fmod Amp = 4 Hz± 2 Hz
1800
D
Fmod F = 25 Hz/sec, Fmod Amp = 4 Hz± 2 Hz
1800
E
Fmod F = 100 Hz/sec, Fmod Amp = 4 Hz± 2 Hz
1800