6.8.1 Load Resistance for Multi Box Models
The load resistance values shown in the tables below represent approximately 80% of maximum
current for the model configuration and phase mode shown in the low voltage range with 100
Vrms programmed.
MODEL
3-Phase
Load Res.
3-Phase
Load Curr
1-Phase
Load Res.
1-Phase
Load Curr.
6000Lx/2
10 ohms
8 amps
4 ohms
24 amps
9000Lx/2
12000Lx/2
5 ohms
16 amps
2 ohms
48 amps
6000Lx/2-HV
7 ohms
10 amps
2.5 ohms
30 amps
9000Lx/2-HV or
12000Lx/2-HV
3.5 ohms
20 amps
1.5 ohms
60 amps
6000Lx/2-EHV
15 ohms
7 amps
4.8 ohms
21 amps
9000Lx/2-EHV or
12000Lx/2-EHV
7.5 ohms
14 amps
2.4 ohms
42 amps
Lx Series - Current Measurement Calibration Currents (2 boxes)
MODEL
3-Phase
Load Res.
3-Phase
Load Curr
1-Phase
Load Res.
1-Phase
Load Curr.
9000Lx/3
5 ohms
16 amps
2 ohms
48 amps
13500Lx/3
18000Lx/3
3.3 ohms
24 amps
1.3 ohms
72 amps
9000Lx/3-HV
3.5 ohms
20 amps
1.5 ohms
60 amps
13500Lx/3-HV
18000Lx/3-HV
2.3 ohms
30 amps
0.83 ohms
90 amps
9000Lx/3-EHV
7.5 ohms
14 amps
2.4 ohms
42 amps
13500Lx/3-EHV
18000Lx/3-EHV
5 ohms
21 amps
1.6 ohms
63 amps
Lx Series - Current Measurement Calibration Currents (3 boxes)
MODEL
3-Phase
Load Res.
3-Phase
Load Curr
1-Phase
Load Res.
1-Phase
Load Curr.
6000Ls/2
7 ohms
10 amps
2.5 ohms
30 amps
9000Ls/2
12000Ls/2
3.5 ohms
20 amps
1.5 ohms
60 amps
6000Lx/2-HV
10 ohms
8 amps
4 ohms
24 amps
9000Ls/2-HV or
12000Ls/2-HV
5 ohms
16 amps
2 ohms
48 amps
6000Ls/2-EHV
15 ohms
7 amps
4.8 ohms
21 amps
9000Ls/2-EHV or
12000Ls/2-EHV
7.5 ohms
14 amps
2.4 ohms
42 amps
Ls Series - Current Measurement Calibration Currents (2 boxes)
MODEL
3-Phase
Load Res.
3-Phase
Load Curr
1-Phase
Load Res.
1-Phase
Load Curr.
9000Ls/3
3.5 ohms
20 amps
1.5 ohms
60 amps
13500Ls/3
18000Ls/3
2.3 ohms
30 amps
0.83 ohms
90 amps
9000Ls/3-HV
5 ohms
16 amps
2 ohms
48 amps
13500Ls/3-HV
18000Ls/3-HV
3.3 ohms
24 amps
1.3 ohms
72 amps
9000Ls/3-EHV
7.5 ohms
14 amps
2.4 ohms
42 amps
13500Ls/3-EHV
18000Ls/3-EHV
5 ohms
21 amps
1.6 ohms
63 amps
Ls Series - Current Measurement Calibration Currents (3 boxes)