Kipp&Zonen COMBILOG Data Logger User Manual

Page 58

background image

The results of the inverse functions are given in radians

(2

π

= 360°). At the function arcsin the value +

2

π

will be

assigned to the Arithmetic Channel for a value

>

1 and the

value –

2

π

will be assigned for a value

<

-1. At the function

arccos the value 0 will be assigned to the Arithmetic
Channel for a value

>

1 and the value will be assigned for a

value

<

-1.

Minimum and maximum of a channel value (min, max)
With this function the minimum and maximum value of a
channel appeared since the last reset has been triggered
off can be determined (“pull-pointer-function”). The result
value can be reset to the actual value of the measured
channel via the bus or at the end of the average interval.

The functions read and write enable the data logger to re-
ceive measured values from other modules connected to
the same bus, resp. to send values to them (master func-
tion)

Note: Logic combinations, e.g. if-then relations, are not yet
possible respectively would require a user-specific software
(upon request).

Hardware Manual COMBILOG 1022