beautypg.com

Kipp&Zonen COMBILOG Data Logger User Manual

Page 123

background image

PROFIBUS - layer 2 commands

DSAP service data to the module (ReqDataUnit) data from the module

(ResDataUnit)

read device identification

0

ident no data

read status information

10

SRD no data

read device information

11

SRD no data

read channel information

12

SRD

read data from a channel

13

SRD

write data to a channel

SRD

response without data

14

SDA

pos./neg. acknowledge

SDN

no response

tare / reset a channel

SRD

response without data

15

SDA

pos./neg. acknowledge

SDN

no response

read, write and tare / reset channels

SRD [[[... [] ] ] ]

...

0

SDA [[[... [] ] ] ]

pos./neg. acknowledge

SDN [[[... [] ] ] ]

no response

Table 11.12 PROFIBUS - layer 2 commands

Note: If more data are in the ReqDataUnit as required, they will
be ignored.

Hardware Manual COMBILOG 1022

123