Kipp&Zonen COMBILOG Data Logger User Manual
Page 162
F.
Accuracy / Resolution / Noise / Linearity / Temp-Drift
Overview measuring ranges
gain
step
gain
factor
voltage meas-
urement
current meas-
urement
resist. measure-
ment
0
1
-
-
20 k
Ω
1
2
±10 V
-
10 k
Ω
2
4
±5 V
-
5 k
Ω
3
8
±2.5 V
25 mA
2.5 k
Ω
4
16
±1.25 V
12.5 mA
1.25 k
Ω
5
32
±625 mV
6.25 mA
625
Ω
6
64
±312.5 mV
3.125 mA
312.5
Ω
7
100
-
-
200
Ω
8
200
±100 mV
1000 µA
-
9
800
±25 mV
250 µA
-
A
3200
±6.25 mV
62.5 µA
-
Hardware Manual COMBILOG 1022
162