Kipp&Zonen COMBILOG Data Logger User Manual
Page 137
x ... “V“:
Universal program
y.yy :
Version
Status information
Register
Type Contents
Length
0500
ro
Module status
2 Byte
0501..0502 ro
Channel status
4 Byte
= hex xy
= 1. Byte
M8..M5 , M4..M1
= hex xy
= 2. Byte
If the bit Mn in the module status is set it indicates that an error
has occurred in the sensor module, the following applies:
M1 = 1: EEPROM - Error
M2 = 1: Flash - Error
M3 = 1: ADC - Error
M4 = 1: Configuration - Error
M5 = 1: No memory card
M6 = 1: RTC - Error
M7...M16 Are currently not occupied
K32..K29 ,
K28..K25
= hex xy
= 1. Byte
K24..K21 ,
K20..K17
= hex xy = 2. Byte
K16..K13 ,
K12..K9
= hex xy = 3. Byte
K8..K5 ,
K4..K1
= hex xy = 4. Byte
If the bit Kn in the channel status is set it indicates that an error
has occurred in channel n.
Hardware Manual COMBILOG 1022