beautypg.com

Resistance measurement, Connection scheme circuit – Kipp&Zonen COMBILOG Data Logger User Manual

Page 42

background image

Resistance Measurement

Figure 5.5 Resistance measurement in 2-wire technique

Figure 5.6 Resistance measurement in 3-wire technique

Figure 5.7 Resistance measurement in 4-wire technique

Hardware Manual COMBILOG 1022

42

Source

IN+

IN-

AGND

U

Rx = U/U0 * R0,

Anschlussschema

Schaltung

Rx

Rx

Messwiderstand

RL

RL

R0

=

U0

Rx = 2 * RL

Connection scheme

Circuit

Source

IN+

IN-

AGND

U1

Rx = (U1/U0-2*U2/U0) * R0,

Anschlussschema

Schaltung

Rx

Rx

Messwiderstand

RL

RL

R0

=

U0

U2

Rx = 0

Connection scheme

Circuit

Source

IN+

IN-

AGND

U2 - U3

Rx = (U2 - U3)/U0 * R0,

Anschlussschema

Schaltung

Rx

Rx

Messwiderstand

RL

RL

R0

=

U0

Rx = 0

Connection scheme

Circuit