Kipp&Zonen COMBILOG Data Logger User Manual
Page 111
Check
sum
request telegram
reply with orderly
performance
reply in
case of er-
ror
Repeat reading the events from pointer 2
with
# aa f cc
> 1 b..b cc
> 0 e cc
NAK
Pointer setting (2) to memory start
with
# aa c cc
ACK
NAK
Pointer setting (2) to a certain date
with
# aa c t..t cc
ACK
NAK
Pointer setting (2) to a position X
with
# aa c x..x cc
ACK
NAK
Sending password
with
# aa P p.p cc
ACK
NAK
write date and time
with
# aa G t..t cc
ACK
NAK
read date and time
with
# aa H cc
> t..t cc
NAK
read measuring rate and averaging interval
with
# aa X cc
> n..n cc
NAK
write measuring rate and averaging interval
with
# aa Y n..n cc
ACK
NAK
delete data memory
with
# aa C.ALL cc
ACK
NAK
read number of events
with
# aa N cc
> a..a cc
NAK
transparent modus on (only at masterfunction)
with
# aa T1 cc
ACK
NAK
transparent modus off (only at masterfunction)
with
# aa T0 cc
ACK
NAK
Table 11.9 Instruction set in the ASCII-Protokoll
Note: In case a password is entered, by means of the configur-
ation program, this password has to be transmitted to the data
logger before memory reading or erasing respectively before
Hardware Manual COMBILOG 1022
111