Kipp&Zonen COMBILOG Data Logger User Manual
Page 113
device identification (v...v)
length = 28 char
ASCII ("Friedrichs")
10 char
ASCII ("COMBILOG")
8 char
ASCII ("xy.yy")
5 char
ASCII ("xy.yy")
5 char
x ... “M“
: monitor program
x … ”A”,“U“: universal program
x ... “T“
: calibration and test program
x ... “S“
: application specific program
y.yy
: version
device information (s...s)
length = 28 char
ASCII
20 char
ASCII
6 char
ASCII
2 char
status information (z...z)
length = 12 char
< channel status >
ASCII
8 Char
< module status >
ASCII
4 Char
K28..K25 …
K8..K5
K4..K1
Byte:
1
2
…
7
8
M12..M9
M8..M5
M4..M1
Byte:
9
10
11
12
If the bit Kn in the channel status is set it indicates that an error
has occurred in channel n. A channel error is given when the
measuring value is outside of the linearisation, e.g. in case of a
sensor break down or of a short circuit of transmission.
Hardware Manual COMBILOG 1022
113