Kipp&Zonen COMBILOG Data Logger User Manual
Page 167
Resistance measurement
range
accuracy
(1)(3)
Resolution /
noise
(1)
linearity
(2)
temperat.
drift
(2)
20 k
Ω
10
Ω
10 k
Ω
5
Ω
5 k
Ω
2.5
Ω
0.003 %
2.5 k
Ω
1.25
Ω
0.05 %
1.25 k
Ω
0.63
Ω
0.01 %
25 ppm/K
625
Ω
0.31
Ω
312.5
Ω
0.15
Ω
0.010 %
5 ppm/K
(4)
200
Ω
0.1
Ω
0.003 %
(1)
values for T=20 °C , (2) values estimated from the
hardware specification
(3)
values correspond to 80 % of the measuring range
(4)
with external drift compensation, reference resistance
with TCR 1
Achievable absolute accuracy with voltage measuring incl. drift correction:
typical
±
3 µV
Achievable accuracy for temperature measurement with Pt 100 resistance
(measuring range -200°...850 °C, temperature range -200°...200 °C, scan
rate 1 s, averaging interval 1 min)
without drift correction:
≤
0.1 °C with ambient temperatures
0 °C...+40 °C
with external drift correction:
typical 0.05°C with ambient temperat-
ures
-40 °C...+85 °C
Reference resistance 100
Ω
, TCR 1
Hardware Manual COMBILOG 1022