2 instruction set in the ascii-protocol – Kipp&Zonen COMBILOG Data Logger User Manual
Page 110
11.6.2
Instruction Set in the ASCII-Protocol
Check
sum
request telegram
reply with orderly
performance
reply in
case of er-
ror
read device identification
with
# aa V cc
> v..v cc
NAK
read device information
with
# aa S cc
> s..s cc
NAK
read status information
with
# aa Z cc
> z..z cc
NAK
read channel information
with
# aa B kk cc
> i..i cc
NAK
read data from one channel
with
# aa R kk cc
> d..d cc
NAK
write data to a channel
with
# aa W kk d..d cc
ACK
NAK
reset / tare a channel
with
# aa D kk cc
ACK
NAK
Read the events from pointer 1
with
# aa E cc
> 1 b..b cc
> 0 e cc
NAK
Repeat reading the events from pointer 1
with
# aa F cc
> 1 b..b cc
> 0 e cc
NAK
Pointer setting (1) to top of memory
with
# aa C cc
ACK
NAK
Pointer setting (1) to a certain date
with
# aa C t..t cc
ACK
NAK
Pointer setting (1) to position x
with
# aa Cx..x cc
ACK
NAK
Read the events from pointer 2
with
# aa e cc
> 1 b..b cc
> 0 e cc
NAK
Hardware Manual COMBILOG 1022
110