If the bit Mn in the module status is set it indicates that an error
has occurred in the data logger. Valid is:
M1 = 1: EEPROM error
M5 = 1: RTC error
M2 = 1: FLASH error
M6
(currently not occupied)
M3 = 1: ADC error
…
M4 = 1: configuration error
M16
(currently not occupied)
channel information (i...i)
length = 32 char
ASCII
1 char
ASCII
20 char
ASCII
1 char
ASCII
1 char
ASCII
1 char
ASCII
6 char
ASCII
1 char
ASCII
1 char
Coding
ASCII "0" Empty Channel (EM)
ASCII "1" Analog Input Channel (AI)
ASCII "2" Arithmetic Channel (AR)
ASCII "3" Digital Output Channel (DO)
ASCII "4" Digital Input Channel (DI)
ASCII "5" Setpoint Channel (VO)
ASCII "6" Alarm Channel (AL)
Coding
Bit 1: tare/reset possible
Bit 2: average value storage configured
Hardware Manual COMBILOG 1022
114