beautypg.com

Reznor SHH (Indoor PreevA) Unit Installation Manual User Manual

Page 39

background image

Form I-PDH/SDH/PEH/SHH/PXH, Page 39

TABLE 24 - Sensor

Data for Johnson A99

Series Temperature

Sensors used in digital

controls - Resistance

VS Temperature

°F

°C

Ohms

-40

-40

613

-31

-35

640

-22

-30

668

-13

-25

697

-4

-20

727

5

-15

758

14

-10

789

23

-5

822

32

-0

855

41

5

889

50

10

924

59

15

960

68

20

997

77

25

1035

86

30

1074

59

35

1113

104

40

1153

113

45

1195

122

50

1237

131

50

1279

140

60

1323

149

65

1368

158

70

1413

167

75

1459

176

80

1506

185

85

1554

194

90

1602

203

95

1652

212

100

1702

221

105

1753

230

110

1804

239

115

1856

248

120

1908

Do not mount the sensor in the ductwork after a split in the supply as that will

cause loss of control in the duct that does not house the sensor.

3. The position of the sensor holder is important. The holder will extend 9-3/16”

(233mm) into the ductwork.
In horizontal ductwork, locate the sensor assembly in the top, middle of the duct

with the sensor probe extending vertically down into the center of the airstream.

In vertical ductwork, locate the sensor assembly in the middle of the side of the

duct that corresponds with the top middle of the discharge outlet.
Turn the holder so that the element will be shielded from direct airflow and will

sense the air temperature as it flows through the holes in the holder.
At the selected location in the ductwork, mark the diamond-shaped hole required

for the sensor holder. Cut the hole no larger than required, approximately 1” x 1”

(25mm x 25mm).

4. The procedure for installing the sensor and attaching the holder depends on

whether the sensor is a capillary or an electrical sensor. Follow the instructions

that apply.

Capillary Sensor (Option AG3 and AG60) - Locate the sensor capillary and run

it out through the hole in the discharge panel of the heater. Determine where the

sensor capillary should enter the box and remove the knockout. Put the capillary

through the hole and secure the bulb to the clip in the holder. Slide the holder into

the ductwork. Using four field-supplied No. 6 sheetmetal screws, attach the box

portion of the holder to the ductwork. Attach the box cover.

Sensor with Wire (Options AG15, AG16, AG58, AG61, AG62, DG5, DG6, D12B,

D12C, D12D, D12E, D12F, and D12G and field-supplied sensor for Option AG40)

- Push the element into the clip in the holder. Determine where the sensor wire

should enter the box and remove the knockout. Slide the holder into the ductwork.

Using four field-supplied No. 6 sheetmetal screws, attach the box portion of the

holder to the ductwork. Attach a field-supplied cable connector to the box, connect

the sensor wire, and attach the box cover.
If sensor is digital, follow the wiring instructions above.
To test the accuracy of the sensor, measure the ohms. Refer to

TABLE 24 (left) to

find the corresponding temperature.

Instructions for Installing Discharge Air Sensor Holder on the Unit -

applies only to Models PDH, SDH, PXH, and SHH without ductwork

1. Installation requires the discharge air sensor holder (FIGURE 32) and the bracket

in

FIGURE 33. (If attached, remove box cover; it will not be used.)

Attach the

bracket

to the box

and the

front of the

heater.

Holder must

be in this

orientation.

Sensor

Element

in Clip

Center post of

a Size 350

Run the capillary

or wire across top

front of the heater.

FIGURE 33 - Discharge Air

Sensor Holder, P/N 115850,

and Bracket, P/N 213612

2. Select a location for the box on the front near the center of the heater. Orientation

of the sensor holder is important. Position the box so that the element will be

sensing the air temperature as it flows through the holes in the holder.