Charge the ear thermometer battery, Spo2, Spo2 frame – Welch Allyn Connex Integrated Wall System - User Manual User Manual
Page 93
•
If the probe is positioned correctly in the ear canal the ExacTemp light flashes. When the
thermometer detects an accurate measurement, the ExacTemp light is continuously on, a
long beep signals the end of the measurement, and the display shows the result.
•
If the probe is positioned incorrectly in the ear canal or is moved during the measuring
process, the ExacTemp light goes out, a sequence of short beeps sounds, and the error
message POS (position error) appears.
7. When you are finished taking the temperature, press the ejector button to eject the used probe
cover.
8. Return the thermometer to the dock.
The LED on the dock flashes while the measurement is being transferred.
After the transfer is complete, the temperature and the temperature scale appear on the Home
tab according to the monitor settings.
Note
Only the latest measurement is transferred to the monitor.
Note
Measurements that have already been transferred to the monitor
cannot be transferred again.
For more information about thermometer functionality, refer to the thermometer manufacturer's
directions for use.
Change the temperature scale on the ear thermometer
To switch from Celsius to Fahrenheit, refer to the thermometer manufacturer's directions for use.
Charge the ear thermometer battery
To charge the battery pack:
•
Place the thermometer in the dock.
•
Make sure that the monitor is connected to AC power.
•
Make sure that the monitor is powered on.
The LED on the dock indicates the charging status of the battery pack:
•
Orange: The battery pack is charging.
•
Green: The battery pack is charged.
•
Not illuminated: The battery pack is not charging.
Note
The battery pack continues to charge while the monitor is in Display power saving
mode.
Note
It is strongly recommended that you use only the Welch Allyn rechargeable battery
pack in the thermometer because the dock cannot charge other batteries.
SpO2
SpO2 and pulse rate monitoring continuously measures saturation level of oxygen in hemoglobin as
well as the pulse rate in a patient through a pulse oximeter.
SpO2 frame
The SpO2 frame displays data and controls used in pulse oximetry measurements.
Directions for use
Patient monitoring 87