Electromagnetic emissions
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
WARNING This equipment/system is intended for
use by healthcare professionals only. This
equipment/ system may cause radio interference or
may disrupt the operation of nearby equipment
a
. It
may be necessary to take mitigation measures, such
as re-orienting or relocating the monitor or shielding
the location.
Complies
a
The monitor contains a 5-GHz orthogonal frequency-division multiplexing transmitter or a 2.4-GHz frequency
hopping spread-spectrum transmitter for the purpose of wireless communication. The radio is operated
according to the requirements of various agencies, including FCC 47 CFR 15.247 and R&TTE Directive (1995/5/
EC). The transmitter is excluded from the EMC requirements of 60601-1-2, but should be considered when
addressing possible interference issues between this and other devices.
Electromagnetic immunity
The monitor is intended for use in the electromagnetic environment specified below. The customer or the
user of the monitor should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
± 1 kV differential
mode
± 2 kV common
mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
>95% dip in 0.5
cycle
60% dip in 5 cycles
30% dip for 25
cycles
>95% dip in 5
seconds
>95% dip in 0.5
cycle
60% dip in 5 cycles
30% dip for 25
cycles
>95% dip in 5
seconds
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the monitor requires continued operation
during power mains interruptions, it is
recommended that the monitor be powered from
an uninterruptible power supply or a battery.
120 Guidance and manufacturer's declaration
Welch Allyn Connex
®
Integrated Wall System