70 percent isopropyl alcohol, 10 percent chlorine bleach solution, Clean the handle cradles – Welch Allyn Connex Integrated Wall System - User Manual User Manual
Page 111: Clean the wall system accessories

70 percent isopropyl alcohol
Wipe the monitor with a clean cloth slightly dampened with 70 percent isopropyl alcohol.
10 percent chlorine bleach solution
1. Wipe the monitor with a clean cloth slightly dampened with a 10 percent bleach and water
solution. Follow the cleaning agent manufacturer's guidelines.
2. Rinse with a clean cloth slightly dampened with water that meets EP and USP quality
standards.
3. Allow the monitor surface to dry for a minimum of 10 minutes before using the monitor.
Clean the handle cradles
The blue handle cradles in the wall system require special attention.
CAUTION Do not use CaviWipes™ to clean the blue handle cradles. This cleaning
agent produces bubbles and liquid during the cleaning process that can enter
openings in the cradles and limit handle performance for as long as 30 minutes after
cleaning.
Clean the handle cradles on a routine basis according to your facility's protocols and
standards or local regulations.
The following agents are compatible with the handle cradles:
•
Sani-Cloth
®
Plus
•
70 percent isopropyl alcohol
•
10 percent chlorine bleach solution
Note
Disinfect according to your facility's protocols and standards or local regulations.
Clean the wall system accessories
The cleaning procedures for wall system accessories vary from the procedure to clean the wall
system itself.
1. Wipe the NIBP hose and any reusable cuffs with a damp cloth moistened in a mild detergent
solution.
2. Wipe the temperature probe with a cloth dampened with alcohol, warm water, or an
appropriately diluted, nonstaining disinfectant solution.
3. Clean the pulse oximetry sensors with a cloth dampened with 70 percent isopropyl alcohol or
10 percent chlorine bleach solution.
CAUTION Never immerse any wall system accessories.
4. Clean the ear themometer according to the manufacturer's directions for use.
5. Clean the physical assessment instrument handles and cords using the same cleaning agents
used on the wall system. Clean on a routine basis according to your facility's protocols and
standards, or local regulations.
6. To clean Welch Allyn 3.5V instrument heads attached to the physical assessment handles,
follow instructions provided in their directions for use.
Directions for use