Minimum alarm volume
Select the minimum alarm volume available. If you
select
High
, then
Medium
and
Low
are not
available to the clinician.
These controls are on the Alarms tab (on the
General tab).
Nurse call threshold
Select the minimum priority alarm that activates a
nurse call relay. If you select
High
, only high-level
alarms activate a nurse call relay.
Audio pause time
Specify the amount of pause time that is added to
the 60-second pause time. When a clinician pauses
an audio alarm tone, the tone is paused for the
combined amount of time.
SpO2 alarm condition delay
Specify the minimum amount of time that an SpO2
alarm condition must be active before audio and
visual signals occur.
SatSeconds is available with Nellcor SpO2 sensors.
If you select 0 seconds or 10 seconds, SatSeconds
is disabled, and it is removed from the SpO2 tab in
the Alarms tab.
SpHb alarm condition delay
Specify the minimum amount of time that an SpHb
alarm condition must be active before audio and
visual signals occur.
4. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify advanced display settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Display tab.
3. Specify settings.
Setting
Action/Description
Display lock
Specify the required period of clinician inactivity
before the touchscreen locks.
Display power saver
Specify the required period of monitor inactivity
before the display turns off.
Clinician interactions, new vital sign
measurements, or alarm conditions automatically
turn on the display.
Device power down
Specify the required period of monitor inactivity
before the monitor turns off.
4. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch Exit.
Directions for use
Advanced settings 125