Specify advanced spo2 settings, Specify advanced sphb settings – Welch Allyn Connex Integrated Wall System - User Manual User Manual
Page 135

clinicians remove the temperature probe from the
well.
Select
Last Site
to set the default to the site
selected for the last measurement.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify advanced SpO2 settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Parameters tab.
3. Touch the SpO2 tab.
4. Specify settings.
Setting
Action/Description
Default view
Select a numeric view or a waveform view as the
primary SpO2 display on the Home tab.
Default response
Select the default speed of response to changes in
SpO2 measurements.
Sweep speed
Select the waveform sweep speed for the SpO2
display in the Home tab.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced tabs and return to the Home tab, touch Exit.
Specify advanced SpHb settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Parameters tab.
3. Touch the SpHb tab.
4. Specify settings.
Setting
Action/Description
Reference
Select arterial or venous as the calibrated
reference source.
Directions for use
Advanced settings 129