Startup, Power – Welch Allyn Connex Integrated Wall System - User Manual User Manual
Page 39

Startup
Power
The power button, located on the front of the monitor, performs the following functions:
•
Powers up the monitor
•
Sets the monitor into Display power saving mode, except when an alarm condition is active
(brief press)
•
Resets the monitor and sets the monitor into Standby mode (press and hold for 6 seconds)
CAUTION Do not use a long press of
to power down the monitor when it is
functioning normally. You will lose patient data and configuration settings.
The LED in the center of the power plug symbol indicates the battery charging status:
•
Green indicates that AC power is present and that the battery is fully charged.
•
Amber indicates that AC power is present and that the battery is charging.
The monitor has distinct power states.
Monitor on
The monitor is operating on battery power or AC power. You can utilize the monitor's features, and
the display is active.
Display power saving
The monitor is operating on battery or AC power, but the display is off to conserve power. A brief
press of the power button sets the monitor into Display power saving mode from the active state.
Settings for this mode can be changed in the Advanced Settings Display tab.
Battery-powered accessories connected to the monitor continue to charge while the monitor is in
this mode and connected to AC power.
Note
The monitor will not enter the Display power saving mode while an alarm condition
is active. In addition, the monitor will exit this mode if an alarm occurs.
33