Maintenance and service, Perform periodic checks, Remove the wall system from the wall – Welch Allyn Connex Integrated Wall System - User Manual User Manual
Page 107

Maintenance and service
Perform periodic checks
Welch Allyn recommends that each facility conduct periodic checks of each monitor.
1. Check the following at least daily:
•
Audio (speaker and piezo beeper tones), especially at startup
•
Fan, especially at startup
•
Touchscreen alignment
•
Date
•
Time
2. Visually inspect the following at least weekly:
•
the monitor for any damage or contamination
•
all cables, cords, and connector ends for damage or contamination
•
all mechanical parts, including covers, for integrity
•
all safety-related labeling for legibility and adhesion to the monitor
•
all accessories (cuffs, tubing, probes, sensors) for wear or damage
•
documentation for current revision of the monitor
3. Visually inspect the following at least monthly:
•
Mounting screws on wall for looseness and wear
Update settings, replace items, or call for service as necessary based on results of visual
inspection. Do not use the monitor if you see any signs of damage. Qualified service personnel
must check any monitor that is damaged for proper operation before putting the monitor back into
use.
CAUTION Wall system components shall be replaced by Welch Allyn service
centers or qualified service personnel.
Remove the wall system from the wall
For maintenance or service activities that require access to the back of the wall system, follow
these steps to remove the wall system from the wall.
1. Touch the Settings tab.
2. Touch the Device tab.
3. Touch Power down.
101